09-0062 response to Container-Quinn Testing Laboratories, Inc. concerning 173.197, 178.603.

## Document text

<<<PAGE 1>>>

U.S. Department of Transportation Pipeline and Hazardous Materials
Safety Administration
1200 New Jersey Ave, S.E.
Washington, D.C. 20590
Mr. Stephen C. Powell
Lab Director
Container-Quinn Testing Laboratories, Inc.
170 Shepard Avenue
Wheeling, IL 60090
Ref. No.: 09-0062
Dear Mr. Powell:
This in response to your March 18,2009 letter regarding the drop test requirements specified in
the Hazardous Materials Regulations (HMR; 49 CFR Parts 17 1-1 80) for a combination
packaging intended for the transportation of regulated medical waste (RMW). The combination
packaging consists of a UN 4H2 plastic container with an inner 3-mil polyethylene bag. You ask
for clarification of the failure criteria in fj 178.603(0(4) and specifically if it is acceptable for the
inner packaging to develop punctures and Yt" to %" holes upon impact if there is no leakage of
the filling substance.
The answer is no. As you correctly noted in your letter, fj 178.603(0(4) states a combination
packaging passes the drop test if there is no damage to the outer packaging likely to adversely
affect the packaging during transport, and there is no leaking of substance from the inner
packaging. It is the opinion of this Office that an inner packaging of a combination packaging
that develops punctures and ?A" to %" holes after the drop test does not meet the requirements in
$178.603(f)(4) and, therefore, does not pass the drop test. The inner packaging's inability to
resist punctures and holes during testing can be indicative of its integrity under transportation
conditions.
Please note that, in accordance with 9 173.197(e)(l), plastic film bags used as inner packagings
for solid RMW must be marked and certified by the manufacturer as having passed the tests for
tear resistance in ASTM D 1922, "Standard for Tear Resistance of Plastic Film and Thin
Sheeting by Pendulum Method," and for impact resistance in ASTM D 1709, "Standard for Test
Methods for Impact Resistance of Plastic Film by the Free-Falling Dart Method." The film bag
must meet an impact resistance of 165 grams and a tear resistance of 480 grams in both the
parallel and perpendicular planes with respect to the length of the bag.
1 hope this information is helpful. If you have further questions, please do not hesitate to contact
this office.
Hattie L. Mitchell
Chief, Regulatory Review and Reinvention
Office of Hazardous Materials Standards

<<<PAGE 2>>>

CONTAINER-QUINN
TESTING LABORATORIES, INC.
A division of H.H. Holmes Testing Laboratories, 1nc:
170 Shepard Avenue Wheeling, IL 60090 knd"&
Phone: 847-537-9470 Fax: 847-537-9098
e-mail: spowell6j)container-quinn.com 3 17% ~03,&c)@)
Te5-6 ny
18 March 2009 by-006z
Mr. Edward T. Mazzullo
Director, Ofice of Hazardous Materials Standards
US DOT I PHMSA (PHH-10)
1200 New Jersey Avenue, SE
East Building, 2" Floor (PH)
Washington, DC 20590
Mr. Mazzullo;
Re. SA-9009004, ID Code +AX
RMW testing of a combination package
After discussion with Anthony Lima, USDOT and at his suggestion, I am requesting an interpretation
and clarification of passlfail criteria in the testing of a Regulated Medical Waste (RIMW) container.
With IWr Lima as an observer, I conducted testing of a 4H2 plastic container, for use with RMW.
Description is as follows: UN 4H2 17-gallon oblong openhead plastic container with sliding transport
cover containing one (1) 3-mil polyethylene red bag. Bag is secured by twisting the top and then
tightly knotting it. Container is secured with cable ties. Test weight is 15.9kg. Drop height was 1.2M
During the testing, the redbag is removed and inspected after each individual drop. 'There were some
holes (114" to W ) and punctures found in the bag after the drops but bag integrity was intact (bag was
lifted completely out of the container and inspected). As per 49CFR, 178.603(9(4), "For a composite
or combination packaging, there is no damage ks the outw packaging likely $0 adverse@ affect safety
during trans~ort, and there is no leakage of t h e fillrng substance f o m the inner packaging;" Under this
section, as the bag was able to contain all product within and the outer container showed no damage,
I considered this to be a pass. Mr Lima raised a concern with the holes in the bag and whether this
compromised the test. This is the area that is under question, whether these holes and punctures are
acceptable and, as the bag does contain all the inner product, if this should be considered a pass in
accordance with the Criteria for Passing the Test as listed above.
Thank you for your assistance in this matter.
Sincerely - --+.
Stephen C. Powell
Lab Director
Container-Quinn Testing Laboratories, Inc.
Cc: Mr. Anthony Lima, US DOTIPHMSA

## Provenance

- Official: Yes
- Source: <https://www.phmsa.dot.gov/sites/phmsa.dot.gov/files/legacy/interpretations/Interpretation%20Files/2009/090062.pdf>
- Source ID: `phmsa`
- SHA-256: `0cbcfd8b9a04807cef1f7f1a6f77a3b38a44aa7e7373d64c7ae6da65b65ef974`
- Retrieved: 2026-08-20T00:59:31.977Z
- Exported: 2026-08-23T10:36:49.579Z
- Document slug: `phmsa-interpretation-09-0062`

### Source metadata

```json
{
  "materialSubtype": "interpretation",
  "interpretationArea": "hazardous_materials",
  "representation": "full_text_from_official_pdf",
  "companies": [
    "Container-Quinn Testing Laboratories, Inc."
  ],
  "individuals": [
    "Mr. Stephen C. Powell"
  ],
  "refIds": [
    "09-0062"
  ],
  "catalogDates": [
    "2009-04-16"
  ],
  "catalogParts": [
    173,
    178
  ],
  "catalogRowCount": 2,
  "sourceRecordUrls": [
    "https://www.phmsa.dot.gov/node/45981"
  ],
  "linkedAttachmentUrls": [],
  "unavailableResponseUrls": [],
  "duplicateRepresentationUrls": [],
  "citedSections": [
    "173.197",
    "173.197(e)(l)",
    "178.603"
  ],
  "catalogPageUrls": [
    "https://www.phmsa.dot.gov/regulations/title49/section/173197",
    "https://www.phmsa.dot.gov/regulations/title49/section/178603"
  ],
  "pdfUrls": [
    "https://www.phmsa.dot.gov/sites/phmsa.dot.gov/files/legacy/interpretations/Interpretation%20Files/2009/090062.pdf"
  ],
  "representations": [
    {
      "viewUrl": "https://www.phmsa.dot.gov/sites/phmsa.dot.gov/files/legacy/interpretations/Interpretation%20Files/2009/090062.pdf",
      "detailUrl": null,
      "detailHtmlPath": null,
      "pdfUrl": "https://www.phmsa.dot.gov/sites/phmsa.dot.gov/files/legacy/interpretations/Interpretation%20Files/2009/090062.pdf",
      "pdfArtifactPath": "data/sources/phmsa-interpretations/09-0062-b49f818dd2.pdf",
      "pdfArtifactSha256": "6ea7b7909eb4defe158cd26baff5f94add4d83eee5dde3a8fe4b5b4f45987f85",
      "extractedTextPath": "data/sources/phmsa-interpretations/09-0062-b49f818dd2.v2.txt",
      "extractedTextSha256": "f02635e2fffdc64ba15c892f9cd50c9351cf3acd2f8d456a89b3ab021e2b7485",
      "pageCount": 2,
      "extractionVersion": 2
    }
  ],
  "caveat": "Interpretations apply regulations to the facts presented and do not create independently enforceable requirements.",
  "jurisdiction": "US"
}
```